1 invasive blood pressure cable & sensor fault pressure not displaying correctly
Possible reasons: host anomaly
Solution: replace the host
Possible reasons: not compatible
Solution: according to the manufacturer's instructions selection
Possible reasons: cable not connected
Solution: connect the cable
Possible reasons: cable fault
Solution: return factory
2 common faults of blood pressure cuff noninvasive blood pressure & accessories display is not correct
Possible reasons: host anomaly
Solution: replace the host
Possible reasons: the testing part sleeves tied up is not correct
Solution: in accordance with the cuff icon tied test site
Possible reasons: cuff selection is not correct
Solution: according to the patient and machine selection right cuff
Possible reasons: the machine is not connected well
Solution: connect the machine end
Possible reasons: the patient jitter
Solution: stop motion
Possible reasons: a short period of time to repeat the test
Solution: the interval of 15 minutes after the test
3 temperature probe and connected to a common failure temperature cannot be displayed
Possible reasons: the machine is not connected well
Solution: connect the machine end
Possible reasons: not compatible
Solution: according to the manufacturers instructions selection
Possible reasons: host anomaly
Solution: replace the host
Possible reasons: probe has been damaged
Solution: return factory
Fault 4 ECG cable and lead wires (1) ECG cannot correctly display
Possible reasons: the machine end or the patient is not connected well
Solution: the ECG cable
Possible reasons: not compatible
Solution: according to the unimed color selection
Possible reasons: the cable is not conductive, wrong polarity
Solution: return factory
(2) waveform burr, unstable
Possible reasons: the patient in the jitter
Solution: stop motion
Possible reasons: next to any large machine interference
Solution: remove interference machine
Possible reasons: Cable anti-interference ability is not good
Solution: return factory
5 oxygen probe & cable fault
(1) no oxygen, no light
Possible reasons: the luminous tube bad
Solution: back to the manufacturer processing
Possible reasons: the luminous tube side cable circuit
Solution: back to the manufacturer processing
Possible reasons: the wrong polarity
Solution: back to the manufacturer processing
Possible reasons: plug loose, not access to the signal with the host
Solution: back to the manufacturer is not compatible with back to the manufacturer processing
Possible reasons: the machine is not connected well
Solution: connect the machine end
(2) without oxygen or waveform
Possible reasons: the receiving tube abnormality
Solution: return to factory repair
Possible reasons: unstable, a light receiving tube cable circuit
Solution: return to factory repair
Possible reasons: not compatible
Solution: choose the right compatible type products
Possible reasons: the leakage calendar harm
Solution: according to the instructions to use
Possible reasons: Probe selection error
Solution: according to the machine instructions to select the correct probe
Possible reasons: test site does not aim at the luminous tube and the receiving tube
Solution: according to the instructions to use
Possible reasons: the patient attention degree is too low to try several times to measure,
Solution: please go to the hospital for accurate diagnosis
Possible reasons: finger tremor or the patient body in motion
Solution: don't move
(3) with oxygen, there is light, but the low oxygen saturation value (40%-60%)
Possible reasons: the wrong polarity
Solution: return to factory repair
(4) testing skin redness
Possible reasons: probe the long-term not loose
Solution: according to the instructions every 2 hours at a time in addition to choose loose, probe models according to the patient's condition
Business problems
1 disposable oversleeve is aseptic packaging? No, not at all. And aseptic packaging is the same packaging, but we have not sterilization.
2 disposable probe shelf-life is? 2 years.
3 repeatability and disposable cuff in the factory test inflation is how many times?
10000 times.
4 neonatal eye is reusable? A single patient use.
5 oxygen line is produced by silicone material? No, by TPU material.
6 can be provided free of charge samples?
Can, but by the customer pays the freight.
7 all products have inventory?
Sufficient for the conventional product inventory.
The 8 united production medical cable for how many years?
10 years
What are the 9 modes of transport?
Shen Tong, SF... Etc.
10 OEM service?
Yes.